Independent-channel LLD on all liquid types is performed via controlled micro-releases of air from the distal end of the tip and subsequent monitoring of pressure changes associated with meniscus (or bubble/foam) interference. A residual volume based on the confirmed meniscus height and labware definition is calculated. Any pipetting discrepancies experienced are addressed via several user-defined error recovery options.
Air-based VVP technology accomplishes robust LLD without the need for more expensive carbon-impregnated (capacitance-based LLD) disposable tips. For the best pipetting results, be sure to use D2-brand ‘performance-certified’ clear disposable tips.
VVP performs continuous millisecond analysis of flow rates for each liquid channel. A disruption in the expected average flow rate, produced either by air entering the channel (i.e. ‘short or no sample’) or a blockage (i.e. clot/clog) triggers an alert.
Diagnosed pipetting errors are addressed automatically ‘on-the-fly’ via user-defined error recovery options.
Independent control of each channel affords the freedom to isolate and disable only the offending channel(s), allowing the method to continue without interruption.
Upon method completion, a final Error Report file is generated.